RIH - CT FOR MASTOIDITIS GE LIGHTSPEED PRO 16 PROTOCOL Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise Index 120kv / smart mA (100-670) / 1.0sec .537:1 , 5…
untitled
…brain reformats parallel to the anterior skull floor (auto-batch off), average mode, auto transferred to PACS Comments: This helical pedi head protocol is the preferred method of scanning when there is any question that the patient cannot maintain a head position that allows the brain to be scanned…
Microsoft Word - pro16 pedi 3d brain.doc
RIH - PEDI 3D HEAD CT GE LIGHTSPEED PRO 16 PROTOCOL Application: For reconstructive surgical planning. Commonly done in cases of Craniosynostosis. Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time…
untitled
RIH - CT PEDI MASTOIDITIS GE LIGHTSPEED PRO 16 PROTOCOL Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise Index 120kv / smart mA (100-670) / 0.8sec 5.37 : 5…
untitled
RIH - 3D HEAD CT GE LIGHTSPEED 16 PROTOCOL Application: For reconstructive surgical planning. Commonly done in cases of Craniosynostosis. Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch…
untitled
RIH - CT FOR MASTOIDITIS GE LIGHTSPEED VCT PROTOCOL Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise Index 120kv / smart mA (100-670) / 1.0sec 0.531:1 , 10… related searches:ct scan protocol
untitled
RIH - CT FOR MASTOIDITIS GE LIGHTSPEED 16 PROTOCOL Position/Landmark Supine head first or feet first Zero at outer canthus of eye. Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise Index 120kv / smart mA (100-440) / 1.0sec 5.37 : 5.37mm…
Pediatric CT Technical Protocols: Pediatric Choose one: GE Lightspeed Pro 16 GE VCT MAS Guidelines for Body CTScans GE Lightspeed QXI Siemens Somatom Sensation 16 NOTE: These protocols require the Adobe Acrobat Reader plug-in. If you do not have the plug-in installed, you can download it now for free. Follow… related searches:ct protocols pediatric | ct protocols | ge ct protocols | siemens ct protocols
untitled
…Scan Delay Archiving to MOD Only prospective recons will be archived to mod as done by the scanner. 2D/3D Technique Used DMPR 5mm x 5mm axial brain reformats in the glabello-meatal plane (autobatch off), average mode, auto transferred to PACS Comments: This helical head protocol is a method of scanning…
untitled
…- if required Scan Delay Hand injection Archiving to MOD 3D Technique Used Comments: brain and bone algorithms are always submitted for a pedi brainct Dose Chart: < 3 years old : 120 mAs 3-6 years old : 180 mAs 7-10 years old : 240 mAs greater than 10 years old : use adult protocol related searches:siemens ct protocols