Head
…;programs" folder. GE Lightspeed Pro 16 Protocols * Brain (3D) * Brain (CTA) * Brain (routine) * Brain (trauma) * Brain (venogram) * Carotid/Brain * Viewing Wand * Stereotactic Gamma Knife * Mastoiditis * Temporal Bones GE Lightspeed QXI Protocols * Brain (routine) * Brain (trauma) * Mastoiditis… related searches:ct protocols | ge ct protocols | protocols
Abdomen/Pelvis
…contrast) * Renal Stone (no IV contrast) * Nissan Fundoplication * Pancreatic Mass * Post EVT * Renal Donor CTA * Renal Mass * 3 Phase Kidney for Tumor Ablation GE Lightspeed QXI Protocols * Abdomen/Pelvis (no IV contrast) * Abdomen/Pelvis IV (routine) * Abdomen/Pelvis IV (acute, trauma) * Abdominal… related searches:ct protocols | ge ct protocols | siemens ct protocols
Chest
CT Technical Protocols: Chest Choose one: GE Lightspeed Pro 16 GE Lightspeed QXI GE VCT GE CTi Protocols GE Lightspeed 16 Siemens Somatom Sensation 16 NOTE: These protocols require the Adobe Acrobat Reader plug-in. If you do not have the plug-in installed, you can download it now for free. Follow… related searches:cta chest protocol | ct protocols | ge ct protocols
untitled
…cta scan delay is the peak from the test bolus injection graph + 10 seconds (because the test bolus has a 8 second scan delay + a 2 second buffer). Recon 1 is a thin soft algorithm for reformats. Sagittal/oblique and coronal reformats, 2.0mm x 2.0mm, average mode using DMPR are routine for this protocol.… related searches:ct protocols neck
Gated/Perfusion Studies
…file should install in your "programs" folder. GE VCT Protocols * Gated Aorta * Gated Coronary CTA * Pulmonary Vein Map * Calcium Scoring Siemens Somatom Sensation 16 * Brain Perfusion * Gated Coronary CTA Back related searches:siemens ct protocols | ge ct protocols | ct protocols
untitled
RIH - UPPER EXTREMITY CTA GE LIGHTSPEED VCT PROTOCOL Position/Landmark Head first or feet first-Supine. The arm should be placed over the patient's head when possible. Zero appropriately Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise…
untitled
RIH - CAROTID AND BRAIN CTA GE LIGHTSPEED VCT PROTOCOL Position/Landmark Supine head first or feet first Zero at sternal notch. Topogram Direction Craniocaudal Scan Type nc brain cta brain axial helical KV / mA / Rotation time (sec) Pitch / Speed (mm/rotation) Noise Index 120 kv / 155 mA / 2 sec 120kv…
untitled
RIH - ABDOMINAL ANGIOGRAM LIGHTSPEED QXI PROTOCOL Indications: Abdominal aortic aneurysm, dissection. This is not to r/o aaa leak. Position/Landmark Head first or feet first-Supine Sternal Notch Topogram Direction Craniocaudal Scan Type Helical KV / mA / Rotation time (sec) Noise Index Pitch / Speed…
untitled
RIH - GATED THORACIC AORTA CTA SIEMENS SOMATOM SENSATION 16 PROTOCOL Position/Landmark Head first or feet first-Supine 2cm superior to shoulders Topogram Direction Craniocaudal Scan Type Helical KV / effective mAs / Rotation time (sec) Pitch 120kv / 350mAs / 0.42 sec .31 : 1 Detector width x Rows = Beam…